Versamed Service Functions; Communication Port; Communication Rate; Localization - GE iVent 201 Series User's Reference Manual

Table of Contents

Advertisement

VERSAMED SERVICE FUNCTIONS

WARNING

COMMUNICATION PORT

COMMUNICATION RATE

LOCALIZATION

WARNING
The service screen is meant for use by VersaMed qualified and trained
personnel only. The technical log book is also meant only for VersaMed
authorized personnel. Do not make changes to the service screen menus
unless you are told to do this by trained VersaMed service technicians.
The iVent
201 has two communication ports, located on the right side of the
TM
rear panel. If you connect a device to the iVent
the communication port.
To configure the communication port:
1.
From the Service screen dial the Control Knob to select the
Communication Port option, then press.
2.
A menu pop-up window appears offering two communication ports;
Com 1, or Com 2. Select the com you need and press the knob to
confirm.
The iVent
201 has a serial port installed for remote communication with other
TM
devices. The default data transfer rate is 115,200 bps. Ordinarily there should be
no reason to change this setting.
To change the communication rate:
1.
From the Service screen dial the Control Knob to select Communication
Rate, then press.
2.
A pop-up window appears offering a range of choices: 1200, 2400, 4800,
9600, 19200, 38400, 57600, and 115200. Make the required selection
and press the knob to confirm.
The Maintenance screen returns with the selected communication rate selected.
The ventilator restarts with the new communication rate.
Localization is used to change the language set for the iVent
Do not access the Localization menu unless you are sure you intend
to change the language the iVent
change the selected language, you change all the screens, menus,
and messages.
201, then you need to configure
TM
201 uses.
TM
186
201.
TM
Remember, if you

Advertisement

Table of Contents