Exclusion – Re-evaluate the patient's clinical status for causes of high respiratory
drive to exclude premature I:E cycling and manage appropriately.
Intervention – Lower the percent peak flow termination (Esens) and/or decrease
the rise time. Use the waveform package to visualize waveforms and prolong the
I-time until a visually detectable inspiratory flow termination pattern appears
prior to pressure release.
Table 9: Recommended Masks
Part Number
Description
Nasal Masks :
All Types (chin strap may be required for mouth
breathers
Respironics Image 3 SE disposable version
Resmed Mirage 2- reusable version (no vent holes)
Koo Medical Equipment- Bluestar Plus NIV Face
M1162031
Mask Kit, Ultra Large (KM-310)
Koo Medical Equipment- Bluestar Plus NIV Face
M1162032
Mask Kit, Large (KM-311)
Koo Medical Equipment- Bluestar Plus NIV Face
M1162033
Mask Kit, Medium (KM-312)
Koo Medical Equipment- Bluestar Plus NIV Face
M1162034
Mask Kit, Small (KM-313)
Koo Medical Equipment- Bluestar Plus NIV Face
M1162035
Mask Kit, Child (KM-314)