INTRODUCTION

This section shows you how to:

CLEANING AND MAINTENANCE

While the iVent
recommended to perform occasional cleaning and basic maintenance
procedures to enhance its lifecycle.
The following chart summarizes cleaning and routine maintenance procedures
and, where appropriate, time intervals.
Part
Ventilator
Air Inlet Filter
Cooling Vents &
Cooling Inlet
Filter
Battery Pack
O
sensor
2
Pneumatic unit
Inlet and outlet
mufflers
Other
accessories
Clean and maintain the iVent
Keep the iVent
201 operable with the help of planned maintenance
TM
Calibrate and perform diagnostic tests on the iVent
201 has been designed to resist damage, staining, and wear, it is
TM
Table 12: Maintenance Procedures and Intervals
Procedure
Wipe exterior clean with
a damp cloth and mild
detergent.
Replace every 500 hours
(or 1 month) of operation
or as necessary.
Clean every 1500 hours
(or 3 months) of use, or
as necessary.
Recharge after every
three months of storage.
Replace every year or as
necessary.
Calibrate every 6 months
Replace every 2 years
Replace every 15,000
hours of operation or
every 4 years, whichever
comes first.
Replace annually or as
necessary
Follow manufacturer instructions.
173
7 Care, Maintenance and Tests
201
TM
201
TM
Comments
Do not attempt to
clean or reuse the air
inlet filter.
Use vacuum device to
clean the vents and
cooling air inlet filter.
Actual life depends on
history of use and
storage.
It is good practice to
calibrate every 3
months
Perform O2 Calibration
and VVT after unit
returns from service
Actual life depends on
history of use and
storage.