5. Press the knob again to save the setting. You are returned to the Parameters
6. Continue through the other parameter options, and choose and change until
NOTE
To change ventilation parameters through the Default Setting Option:
See page 105 Restore Default Settings, for instructions
CAUTION
Selecting Restore Defaults resets ALL values to their default state, including
the Ventilation mode (SIMV Volume Control). For more information about
restoring defaults. See page 105 Restore Default Settings
THE SELECTION INTERFACE
Most of the quantitative parameters are presented in the form of an oval dial with
black numerals – in the iVent
teal. Turning the Control Knob increases or decreases the selected value as
indicated by the black numerals inside the oval. A dark blue ribbon indicator
moves around the periphery of the oval to show the scale of the selected value.
Recommended ranges are outlined in green, while values outside the
recommended ranges are outlined in red. As a precaution, if a setting you are
inserting exceeds or fails to meet recommended settings for the patient weight, a
bright yellow Caution flag appears, as shown in Figure 34 below.
Figure 34: Example of Respiratory Rate exceeding the Recommended Value
Other parameter settings are adjusted on a slider gauge.
screen.
you have set all the values you want. Once you have finished, select
Accept at the bottom of the screen. Press the knob and your new
selections are saved.
To abort the selection process and return the parameter to the previously
chosen value, press the Clear button
66