Warning
Mode of Failure Priority Probable Cause Criterion for
Set Count or
Time of
Condition
Check Sensor
Sensor providing
unreliable
Immediate
information -
ventilator enters
Open Loop mode
SpO2 Patient
The iVent
Disconnect
identifies that the
SpO2 sensor is
connected but
that no signal is
received from it
SpO2 Reading
Low signals from
Failed
the SpO2 sensor
SpO2 Sensor
No signal is
disconnect
received from
the SpO2 sensor
Tube
Tubing
Disconnect
disconnected at
ventilator outlet
Immediate
V
not
Set tidal volume
T
delivered
has not been
(Low V
)
delivered
T
4 breaths
delay (10 with
setting
change)
High
Soiled sensor
Excessive water
in sensor
Sensor tubing
disconnected or
broken
Mid
The patient is
disconnected
from the
oximeter cable
Mid
A bad connection
between the
patient and the
SpO2 sensor, or
SpO2 connectors
are not firmly
connected.
Mid
The SpO2 sensor
is not connected
to the iVent
.
although the
patient might be
connected to the
sensor, the
power to the
sensor is
disconnected, or
if the cable is
damaged
High-
Disconnection of
Mid
tubing
Mid
I time set too low
Rate set too high
Blue tube
disconnected
Pressure limit set
too low
Clinical situation
165
Corrective Action
Clearing Alarm
Normal sensor
Replace patient
reading
circuit
Remove unit for
inspection and
servicing by
qualified service
personnel
When the
Check the patient
patient is
condition.
reconnected to
Reconnect the
the SpO2 sensor
patient to the
for 2 breaths
SpO2 sensor.
The signals
Check if the
from the SpO2
sensor is placed
sensor return to
correctly on the
normal values
patient; try to find
a better place for
connecting the
patient.
Reconnect the
patient to the
SpO2 sensor
When a signal
Connect the SpO2
from the SpO2
sensor to the
sensor is
iVent
received by the
iVent.
Reconnection of
Reconnect tubing
tubes detected
Measured
Increase pressure
Exhale Volume
limit and/ or
is higher than
inspiratory time if
set volume
appropriate
minus 15% for
more than 2
Reconnect blue
breaths
tube
6 Alarms