CIRCUIT RESISTANCE
OTHER CONNECTIONS
HME
CAUTION
NOTE
MDI
USING THE IVENT
HUMIDIFICATION (1.4 STEPPER BASED MODELS
ONLY)
It is important to check the inspiratory and expiratory resistance specification of
the patient circuits used with the iVent
exceed the following limits when adding attachments or other components or
subassemblies to the breathing circuit:
O.6 KPA (6cmH2O) at 60 L/min for adult patients.
O.5 KPA (5cmH2O) at 30 L/min for pediatric patients.
If required, a Heat and Moisture Exchanger may be placed between the flow
sensor and the patient connection.
A contaminated HME filter can impede the ventilator's Patient Disconnect
alarm detection. Do not use any filter that appears to be discolored or filled
with water.
Adding attachments, components or sub-assemblies to the iVent
breathing circuit may increase the pressure gradient.
To administer medication, install an adapter for MDI treatment between the flow
sensor and the patient connection.
TM
201 WITH HEATED
The following procedures explain how to connect the iVent
humidifier.
To use the iVent
201 with a heated humidifier:
TM
1.
Disconnect the Inspiratory Limb from the ventilator outlet of the iVent201
and reconnect it to the humidifier outlet.
2.
Remove the one-way valve from the Patient Wye and reconnect the tube.
201 ventilator to ensure they do not
TM
44
201
TM
201 to a heated
TM