Setting the Apnea Time Alarm

1.
2.
3.
Figure 136: The Pressure Alarm Slider Selected
Press the Control Knob to change the alarm settings for the Pressure.
The top of the slider is selected bright blue. Dial the knob to adjust the alarm
for the upper limit in cmH2O up or down. Note that if you attempt to adjust
the upper limit of the alarm below the currently-selected lower limit, you
push the lower limit down.
Press the Control Knob to confirm your selection.
The bottom of the slider is selected bright blue. Turn the dial to adjust the
alarm for the lower limit in cmH2O. Note that adjusting the lower limit
higher than the currently-selected higher limit moves the higher limit up.
Press the Control Knob to confirm the settings.
The Alarm Setting screen is re-displayed. You can dial and select an
additional alarm setting to adjust, or dial and select Accept to accept all the
alarm settings and return to the Main screen.
Select the Menu option in the Main screen to view the Main Menu.
Select the first item on the Main Menu, Alarm Settings.
In the Alarm Settings screen, dial the Control Knob until Apnea Time is
selected in relief (Figure 137).
149
6 Alarms