Adjusting Flow Termination (Esens) - GE iVent 201 Series User's Reference Manual

Table of Contents

Advertisement

In Adaptive Bi-Level ventilation mode the rise time can be adjusted between 0.1
and 1.5 seconds (Figure 70).
Figure 70: The Rise Time pop up window in Adaptive Bi-level ventilation mode

ADJUSTING FLOW TERMINATION (ESENS)

The iVent
detection to an adjustable percentage of the peak flow. The flow termination
(exhale sensitivity) can be set between 10 to 90 percentages of the peak flow.
To adjust the flow termination:
Figure 69: The Rise Time Pop Up Window
3. Turn the Control Knob to select the required value.
4. Press the Control Knob to confirm and accept the settings.
201 terminates the pressure support breath delivery after flow drop
TM
1.
Select the Esens option in the lower portion of the Main screen (Figure
71):
84

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ivent 201 1.4

Table of Contents