The ranges for Adaptive Bi-Level parameters are as follows:
Rate
FiO
2
Rise Time
IPAP
EPAP
Esens
I-Time
NOTE

ADAPTIVE BI-LEVEL WINDOW

This is what the iVent
Figure 127: Adaptive Bi-Level Mode Parameters
Table 8: Adaptive Bi-Level Parameters' Range
1 to 80 bpm
21% to 100%
0.1 to 1.5 sec
Note: Rise time cannot exceed inspiration time.
7 to 60 cmH
cmH
O
2
Note: IPAP cannot be set below P-Low +2.
0 to IPAP minus 2 or Maximum 30 cmH
10% to 90%
0.2 to 3 seconds; with CAUTION indication at 2.0
seconds or when rate setting I:E ratio reaches 1:4
NOTE: If I-Time flashes, inspiration is terminated
when the set I-Time is reached. When breath
termination occurs due to decreased flow, the I-
Time display is steady.
Apnea Back Up ventilation operates in the same manner as in other
ventilation modes. See Appendix E for more details.
201 Adaptive Bi-Level mode screen looks like (Figure 128):