2: Prerequisites
Location Requirements
The cabinet should be located away from traffic patterns, doors, fans, ventilation
registers, fume hoods and any other air-handling devices that could disrupt its
airflow patterns. All windows in the room should remain closed. Figure 2-1 shows
the preferred location for the cabinet.
Figure 2-1
Preferred location
Air register blocked or redirected
to prevent cabinet disruption
Alternate location
Main Disconnect Device
(Power cord plug in back
of the electronics
module).
Do not position the unit so that it is difficult to operate the main disconnect
device.
Ne placez pas l'appareil de sorte qu'il est difficile de faire fonctionner le
dispositif principal de déconnexion.
9
Need help?
Do you have a question about the Logic+ 30348 Series and is the answer not in the manual?
Questions and answers