Wash hands and arms thoroughly with germicidal soap
Wear appropriate personnel protective equipment (PPE)
Wipe-Down
Raise the sash to its full open position (approximately 21.75 inches or 552
mm). Mute the alarm by pressing
Wipe down the interior surfaces of the cabinet with 70% ethanol, or a
suitable disinfectant, and allowed to dry
Loading Materials and Equipment
Only load the materials required for the procedure. Do not overload the
cabinet
Do not obstruct the front, side, or rear return air grilles
Large objects should not be placed close together
Slowly close the sash until it is in the correct operating position
After loading the cabinet, wait two to three minutes to purge airborne
contaminants from the work area
Work Techniques
Keep all materials at least 4 inches (100 mm) inside from the sash, and
perform all contaminated operations as far to the rear of the work area as
possible
Segregate all clean and contaminated materials in the work area
Arrange materials to minimize the movement of contaminated materials into
clean areas
Keep all discarded contaminated material to the rear of the work area
Avoid moving materials or the operator's hands and arms through the front
access opening during use
Avoid the use of an open flame. Use disposable labware or an electric
incinerator as alternatives
Use proper aseptic technique
Avoid using techniques or procedures that disrupt the airflow patterns of the
cabinet
If there is a spill or splatter during use, all objects in the cabinet should be
surface decontaminated before removal. Thoroughly disinfect the working
area of the cabinet WHILE IT IS STILL IN OPERATION, to prevent the
release of contaminants from the cabinet