15.3 Technical data

Ambient conditions
Storage and transport in original packaging
Storage and transport without packaging
Long-term storage with/without packaging
Operation
Charging the battery
Product
Reference number
Mobility grade according to MOBIS
Maximum body weight
Structural height
Protection rating
Range of Bluetooth connection to PC
Range of Bluetooth connection to remote control
Service life of the prosthesis
Max. insertion depth of the tube adapter
Weight of the prosthesis without Protector
Prosthesis battery
Battery type
Charging cycles (charging and discharging cycles)
after which at least 80% of the original battery capacity
remains available
Charging time until battery is fully charged
Product behaviour during the charging process
Operating time of prosthesis with fully charged battery
Battery charger
Reference number
Storage and transport in original packaging
Storage and transport without packaging
Operation
Input voltage
94 | Ottobock
-25 °C/-13 °F to +70 °C/+158 °F
-25 °C/-13 °F to +70 °C/+158 °F
max. 93% relative humidity, non-condensing
-20 °C/-4 °F to +50 °C/+122 °F
max. 93% relative humidity, non-condensing
-10 °C/+14 °F to +60 °C/+140 °F
max. 93% relative humidity, non-condensing
+10 °C/+50 °F to +45 °C/+113 °F
3C98-3/3C88-3
2 to 4
136 kg/200 lb
max. 240 mm/9.44 inch
IP22
max. 10 m/32 ft
max. 10 m/32 ft
This component was tested for three million load cycles
in accordance with ISO 10328. Depending on the activ­
ity level, this corresponds to a service life of three to five
years.
55 mm/2.17 inch
approx. 1230 g/43.39 oz
Li-Ion
500
7 hours
The product is non-functional
1 day with average use
4E50*
-25 °C/-13 °F to +70 °C/+158 °F
-25 °C/-13 °F to +70 °C/+158 °F
max. 93 % relative humidity, non-condensing
0 °C/+32 °F to +40 °C/+104 °F
max. 93 % relative humidity, non-condensing
12 V
C-Leg 3C98-3/3C88-3