17. Fault messages

84.
85.
86.
Over travelled / Emergency stop
Faulty Emergency Stop Loom.
The emergency stop has two parts a hardware stop circuit and software stop circuit.
The hardware stop circuit and overtravel circuit both use the same signal to trigger a
relay to stop the stand. So if the hardware wire of the Emergency stop loom is
damaged/faulty it will activate the same relay. The software detects this and then
checks to see if it has received a software emergency stop signal.
Number 022-780
Issue Number: 02
"Software Timeout" – The
software has previously
crashed
"Software Timeout Over
travel detected" – The
hard limit switch has been
hit
"Motor 0: overtemp" – The
temperature sensor has
detected that the motor is
over temperature.