Connection To 1482 Inductance Standard; Figure 3-20: Connecting 1693 To 6-Terminal 1482; Figure 3-21: Direct Connection Of 1693 And 1482 - IET Labs, Inc. RLC Digibridge 1693 User And Service Manual

Table of Contents

Advertisement

3.2.11 Connection to 1482
Inductance Standard
Two methods that can be used when calibrating a 1482
Inductance standard is a direct method and a transfer
method, both are described below.
3.2.11.1 Transfer Methods
6-Terminal Transfer method: When lead inductance is
not sufficiently constant, its effect can be eliminated
by another method of comparison. The two six termi-
nal indictors can be connected in series to the bridge.
A first measurement is made with the short of one
inductor in the L position while that of the other is in
the Lo position on the reference terminals. A second
measurement is then made with both links moved to
the opposite terminals. The measured difference is
the difference between the two inductance changes
at the reference terminals and is independent of the
internal lead inductances because they appear in both
measurements and cancel in the difference.
The 1693 Digibridge is connected to the 1482-A
Inductance Standard as illustrated in Figure 3-20. A
14-guage piece of bus wire is used to connect between
the high terminals of both known and unknown stan-
dards. PH and IH are connected to the low side of the
known and PL and IL are connected to the low side
of the unknown. Shorting bars should short between
low and guard terminals for each standard.
1693 RLC DigiBridge
R
L
C
G
Z
Y
Q
D
R
X
B
ANG
.10000
948.8
uF
nF
pF
mH
H
uS
mS
S
k
DEG
PPM
I
P
P
I
L
L
H
H
IL
PL
14 gauge bus wire
GND
LOW
HIGH
100 µH
1482-B STANDARD INDUCTOR
L
L
0
formerly made by GenRad

Figure 3-20: Connecting 1693 to 6-terminal 1482

Operation
EXTERNAL
BIAS
BIAS
TM
ON
On
Off
GO
CONST I.
NO
CONST V.
GO
RANGE
HELD
NOT 1kHz
START
RATIO
REMOTE
On
R/Q
L/Q
C/D
C/R
CONTROL
POWER
SER
PAR
Off
IH
PH
GND
LOW
HIGH
100 µH
1482-B STANDARD INDUCTOR
L
L
0
formerly made by GenRad
3-terminal transfer method: Two measurements are
performed with a 14 gauge bus wire moved so that
the inductance of the DUT is made and then the bus
wire is moved so that the measurement of the stan-
dard is made. The formula below can then be used
to determine the value of DUT.
Known Value of Std
Actual Value DUT =
Measured Value of Std
See Figure 3-22 for the 3-terminal transfer method.
Ratio measurements of 2:1, made on a 1689, on the
same range and using the delta% readout, typically
have errors totaling less than 20 ppm. This allows
comparisons of inductors of intermediate value (those
values starting with a 2 or 5) to be compared against
the even decade values with negligible added error.
3.2.11.2 Direct Measurement
Method
The 1482 Inductor can be directly measured on the
Digibridge. A combination of the 1689-9602 BNC to
BNC cable, two BNC T adapters and Pomona 1894
adaptors are used for connection. Open and Short
compensation should be performed. Spacing should
be maintained when performing open compensation.
To minimize noise the high leads are connected to the
low terminal of the 1482 and the low terminals to the
high terminal of the 1482.
1693 RLC DigiBridge
R
L
C
G
Z
Y
Q
D
R
X
B
1.0000
484.1
uF
nF
pF
mH
H
uS
mS
S
k
DEG
PPM
I
P
P
I
L
L
H
H
IL
PL
IH
PH
GND
LOW
HIGH
1 mH
1482-E STANDARD INDUCTOR
formerly made by GenRad

Figure 3-21: Direct connection of 1693 and 1482

1693 RLC Digibridge
*Measured Value of DUT
EXTERNAL
BIAS
BIAS
TM
ON
ANG
CONST I.
CONST V.
RANGE
HELD
NOT 1kHz
START
RATIO
REMOTE
On
R/Q
L/Q
C/D
C/R
CONTROL
POWER
SER
PAR
Off
On
Off
GO
NO
GO
27

Advertisement

Table of Contents
loading

Related Products for IET Labs, Inc. RLC Digibridge 1693

Table of Contents