IET Labs, Inc. RLC Digibridge 1693 User And Service Manual page 106

Table of Contents

Advertisement

1693 RLC Digibridge
Value
HOLD
Hold range
1
Hold range 1
2
Hold range 2
3
Hold range 3
4
Hold range 4
AUTo
Autorange
Equivalent legacy command:
R0, R1, R2, R3, R4, R5
Equivalent key:
([1] or [2] or [3] or [4]) [=] [SHIFT] [SPECIAL] [1]
:CONFigure:DELay
Syntax:
:CONFigure:DELay <numeric>
Description:
Sets the trigger delay time.
Parameter:
Value
<numeric>
Delay time real value in ms
Equivalent legacy command:
J<numeric>;
Equivalent key:
(delay in ms) [=] [SHIFT] [SPECIAL] [4]
:CONFigure:ITFactor
Syntax:
:CONFigure:ITFactor <numeric>
Description:
Set the integration-time multiplying factor. Value
<numeric> multiplies MEDIUM and FAST integra-
tion times.
Parameter:
Value
<numeric>
Integration-Time Multiplying Factor real value
Equivalent legacy command:
I<numeric>;
Equivalent key:
(IT factor in m) [=] [SHIFT] [SPECIAL] [5]
82
Description
Description
Description
:CONFigure:FREQuency
Syntax:
:CONFigure:FREQuency <numeric>
Description:
Sets the frequency used for measurement signal.
Parameter:
Value
<numeric>
Frequency real value in kHz
Equivalent legacy command:
F<numeric>;
Equivalent key:
(frequency) [=] [SHIFT] [FREQUENCY]
:CONFigure:VOLTage:CONStant
Syntax:
:CONFigure:VOLTage:CONStant {OFF|ON}
Description:
Set the measurement signal voltage.
Parameter:
Value
OFF
Set constant voltage off
ON
Set constant voltage on
Equivalent legacy command:
Y0, Y1
Equivalent key:
[1] [=] [SHIFT] [CONST V]
:CONFigure:VOLTage:LEVel
Syntax:
:CONFigure:VOLTage:LEVel <numeric>
Description:
Set the measurement signal voltage level.
Parameter:
Value
Set measurement voltage level to real value (in
<numeric>
volts)
Equivalent legacy command:
V<numeric>;
Equivalent key:
(voltage) [=] [SHIFT] [VOLTAGE]