IET Labs, Inc. RLC Digibridge 1693 User And Service Manual page 107

Table of Contents

Advertisement

:CONFigure:CURRent
Syntax:
:CONFigure:CURRent <numeric>
Description:
Set the measurement constant current level.
Parameter:
Value
<numeric>
Set measurement current to real value (in mA)
Equivalent legacy command:
i<numeric>;
Equivalent key:
(current) [=] [SHIFT] [CONST I]
:CONFigure:BIAS
Syntax:
:CONFigure:BIAS {OFF|ON|0|1}
Description:
Enables or disables DC bias.
Parameter:
Value
OFF or 0
Disables the DC bias
On or 1
Enables the DC bias
Equivalent legacy command:
U0, U1
Equivalent key:
([0] or [1]) [=] [SHIFT] [SPECIAL] [INT BIAS]
:CONFigure:ROUTine
Syntax:
:CONFigure:ROUTine {NORMal|REVersal|SHORt
ing|REVShortint|QUICk|QSHorting}
Description:
Set the measurement routine.
Operation
Description
Description
Parameter:
Value
NORMal
Normal routine (highest speed)
Signal reversal (use for low frequency
REVersal
hum rejection)
Shorting between measurements
(used for reducing transient-recovery
SHORting
time within measuring circuit, when
measuring with bias; NOT FOR
DISCHARGING THE DUT)
Signal reversal and shorting between
REVShorting
measurements
QUICk
Quick-acquisition feature
QSHorting
Quick-acquisition & shorting
Equivalent legacy command:
Q0, Q1, Q2, Q3, Q4, Q5
Equivalent key:
none
:CONFigure:NOMinal
Syntax:
:CONFigure:NOMinal <numeric>
Description:
Set the nominal real value <numeric> in ohms,
henries, or farads.
Parameter:
Value
<numeric>
Nominal real value
Equivalent legacy command:
N<numeric>;
Equivalent key:
nominal value + [=] [SHIFT] [NOMINAL]
:CONFigure:BLANking
Syntax:
:CONFigure:BLANKing <a numeric> <b numeric>
Description:
Set display digit blanking. Eliminate <a numeric>
digits from RLC and <b numeric> digits from QDR
displays on Digibridge front
1693 RLC Digibridge
Description
Description
83

Advertisement

Table of Contents
loading

Related Products for IET Labs, Inc. RLC Digibridge 1693

Table of Contents