2.1 Unpacking and Inspection
This instrument was carefully inspected before ship-
ment. It should be in proper electrical and mechanical
order upon receipt.
An OPERATING GUIDE is attached to the instru-
ment to provide ready reference to specifications and
operating information.
If the shipping carton is damaged, inspect the instru-
ment for damage (scratches, dents, broken parts, etc).
If the instruments is damaged or fails to meet speci-
fications, notify the carrier and IET Labs for RMA
number. Retain the shipping carton and the packing
material for carrier's inspection.
2.2 Dimensions
Width = 440 mm (17.3")
Height = 147 mm (5.8")
Depth = 377 mm (14.8")
8
Chapter 2
INSTALLATION
2.3 Power Line Connection
This instrument uses a standard IEC input module.
The input fuse is located in the drawer above the
input power module. If it has to be replaced, be sure
to use a "slow blow" fuse, 250 V, 6/10 A, 5x20 mm.
See Figure 2-1.
Figure 2-1: Input power module with a drawer
2.4 Line-Voltage Regulation
The accuracy of measurements accomplished with
precision electronic test equipment operated from
ac line sources can often be seriously degraded by
fluctuations in primary input power. Line-voltage
variations of ±15% are commonly encountered, even
in laboratory environments. Although most modern
electronic instruments incorporate some degree of
regulation, possible power-source problems should
be considered for every instrumentation setup. The
for the input fuse
Installation