IET Labs, Inc. RLC Digibridge 1693 User And Service Manual page 113

Table of Contents

Advertisement

:MEASure
Syntax:
:MEASure
Description:
Start a measurement upon read (like pressing the
start switch).
Equivalent legacy command:
G0
Equivalent key:
[START] key
*CLS
Syntax:
*CLS
Description:
Clears the following:
• Error Queue
• Status Byte Register
• Standard Event Status Register
• Operation Status Event Register
• Questionable Status Event Register (No Query)
• Resets to Turn-On Conditions (legacy P2
command)
Equivalent legacy command:
P2
Equivalent key:
none
*ESE
Syntax:
*ESE <numeric>
*ESE?
Description:
Sets the value of the Standard Event Status Enable
Register to specified numeric value. *ESE? reads
the current value of the Standard Event Status
Enable Register without clearing the register value.
Operation
Parameter:
Value
0 to 255
Numeric value range
Equivalent legacy command:
none
Equivalent key:
none
*ESR?
Syntax:
*ESR?
Description:
Reads out the value of the Standard Event Status
Register. Executing this command clears the
Standard Event Status Register value.
Equivalent legacy command:
none
Equivalent key:
none
*IDN?
Syntax:
*IDN?
Description:
Reads out the Digibridge instrument informa-
tion (manufacturer, model number, serial number,
FPGA version number, firmware version number)
of the Digibridge.
Example: "IET Labs, 1693M, 20110228, 1.0, 1.0"
Equivalent legacy command:
none
Equivalent key:
none
*OPC
Syntax:
*OPC
1693 RLC Digibridge
Description
89

Advertisement