Internal Fuse Replacement - IET Labs, Inc. RLC Digibridge 1693 User And Service Manual

Table of Contents

Advertisement

5.7.3 Internal Fuse Replacement

WARNING
Take appropriate safety precautions
when servicing the unit.
If at power-up, there is a failure display of
555 5:XXXX (see Self Check paragraph 3.13.1 and
Trouble Analysis paragraph 3.7.1) replacement of the
capacitor discharge fuse may be necessary.
Service and Maintenance
Figure 5-11: Location of internal fuse
1. Remove the top and bottom covers.
2. This fuse is located in a socket on the main
board and behind the power switch. Unplug
the fuse (using needle nose pliers) and
replace.
3. Install top and bottom covers.