1693 RLC Digibridge
Parameter:
<a numeric>
Value
0 to 5
Number of digits to eliminate from RLC display
<b numeric>
Value
0 to 4
Number of digits to eliminate from QDR display
Equivalent legacy command:
O<a numeric><b numeric>
Equivalent key:
none
:CONFigure:BIN:NUM
Syntax:
:CONFigure:BIN:NUM <numeric>
Description:
Set the limit entry bin number.
Parameter:
Value
<numeric>
Bin #, range 0 to 15
Equivalent legacy command:
B<numeric>
Equivalent key:
none
:CONFigure:BIN:HIGH
Syntax:
:CONFigure:BIN:HIGH <numeric>
Description:
Set the limit entry bin high limit (in %).
Parameter:
Value
<numeric>
Real value high limit in %
Equivalent legacy command:
Hh;
Equivalent key:
none
84
Description
Description
Description
Description
:CONFigure:BIN:LOW
Syntax:
:CONFigure:BIN:LOW <numeric>
Description:
Set the limit entry bin low limit (in %).
Parameter:
Value
<numeric>
Real value low limit in %
Equivalent legacy command:
L<numeric>;
Equivalent key:
none
:CONFigure:BINSummary
Syntax:
:CONFigure:BINSummary
{DISable|ENABle|RESet}
Description:
Enables or disables bin summary output.
Parameter:
Value
DISable
Disable output
ENABle
Enable output
RESet
Reset bin summary to 0
Equivalent legacy command:
E0, E1, E2
Equivalent key:
none
:CONFigure:OUTPut
Syntax:
:CONFigure:OUTPut{NONe|BIN|QDR|QBIN|RL
C|RBIN|RQDR|RQBin|RQBFail|RQBPass}
Description:
Selects the data that is output.
Description