1693 RLC Digibridge
Command
Parameter(s) Description
OFF
CONStant
ON
LEVel
<numeric>
Sets constant current
CURRent
<numeric>
Enables or disables DC bias
BIAS
OFF
ON
Sets the measurement routine
NORMal
REVersal
SHORting
ROUTine
REVShorting
QUICk
QSHorting
Sets the nominal value
NOMinal
<numeric>
Front panel display digit blanking
BLANking
<a> <b>
BIN:
Sets the limit entries for binning
NUM
<numeric>
HIGH
<numeric>
LOW
<numeric>
Enables or disables bin summary output
DISable
BINSummary
ENABle
RESet
78
Legacy
Constant
Y0
voltage off
Constant
Y1
voltage on
Real value in
V<numeric>;
volts
Real value in
i<numeric>;
mA
Off
U0
On
U1
Normal routine
Q0
(highest speed)
Signal reversal
(for low freq
Q1
hum rejection)
Shorting
between
Q2
measurements
Signal reversal
Q3
and shorting
Quick-
acquisition
Q4
feature
Quick-
acquisition &
Q5
shorting
Real value in
ohms, henries,
N<numeric>;
farads
Eliminate <a>
digits from RLC
O<a><b>
and <b> digits
from QDR
Bin #
B<numeric>
Real value high
H<numeric>;
limit in %
Real value low
L<numeric>;
limit in %
Disable output
E0
Enable output
E1
Reset bin
E2
summary to 0
Command
Parameter(s) Description
Sets the data to output via GPIB / RS-232
NONe
BIN
QDR
QBIN
RLC
OUTPut
RBIN
RQDR
RQBin
RQBFail
RQBPass
Selects GPIB / RS-232 output data format
ASCii
DATaformat
BINary
SYSTem:
Enables or disables the manual start switch
ENABle
SWITch
DISable
Enables or disables the front panel keypad
ENABle
KEYPad
DISable
RESet
RESTart
Zero calibration
DISable
OPEN
CALibration
SHORt
COMMunicate:
GPIB:
ADDRess?
Enables or disables the GPIB interface
ENABle
CONFigure
DISable
SERial:
Legacy
None
X0
Bin #
X1
QDR
X2
QDR, Bin #
X3
RLC
X4
RLC, Bin #
X5
RLC, QDR
X6
RLC, QDR,
X7
Bin #
RLC, QDR, Bin
X8
# on FAIL only
Same on FAIL,
X9
Bin # on PASS
Normal ASCII
x0
(not compacted)
Compacted
x1
binary format
W0
Enable switch
Disable switch
W1
Unlock (enable
K0
keypad)
Lock (disable
K1
keypad)
Resets settings
to turn-on
P2
conditions
Restarts and
reruns power-on
P3
diagnostics
Disable
Z0
Perform
open circuit
Z1
calibration
Perform
short circuit
Z2
calibration
Reads the
instrument GPIB
address
Enable GPIB
interface
Disable GPIB
interface (if
using RS-232)