IET Labs, Inc. RLC Digibridge 1693 User And Service Manual page 117

Table of Contents

Advertisement

Character
Sequence
Status
1
2
Format
Parameter
3
4
Format
Units
5..7
7
Format
Sign
8
Number
9..15
16
17
* If measurement is overrange (display goes blank), 999999 is output in this field.
** Two leading zeros will precede the decimal point on values less than 1. Two spaces will precede the value on values greater
than 1.
Character
Sequence
Pass/fail
1
2
Format
3
Label
4
5
6
Format
7,8
Bin Number
9
10
Operation
Allowed
Purpose
Characters
(space)
O
I
(space)
Q
D
R
X
B
A
(space)
(2 space)Ω
(space)kΩ
(2 space)S
(space)mS
(space)µS
deg
ppm
(3 space)kO
(space)
(space)
-
0123456789.
(space)**
(CR)
(LF)
Table 3-22: QDR-Value data output format
Allowed
Purpose
Characters
(space)
F
(space)
B
I
N
(space)
123456789
(CR)
(LF)
Table 3-23: BIN-Number data output format
Meaning
Normal Operation
Overrange held (reduced accuracy)*
Invalid measurement due to signal overload
Quality factor
Dissipation factor
Resistance
Reactance
Susceptance
Angle
Ohms
Kiloohms
Siemens
Millisiemens
Microsiemens
Degrees
Part per million (for D, Q or Angle)
Dimensionless (for Q or D)
Positive Value
Negative Value
Measured number, right justified in format field, like the QDR
display except the zero before the decimal point is explicitly
provided and this number can be as long as six characters. *
The standard "carriage return" and "line feed" characters;
end of string.
Meaning
Go (bins 01 to 13)
No-Go (bins 00 or 14)
The word "BIN"
Bin number assignment, 00 to 14
The standard "carriage return" and "line feed" characters;
end of string.
1693 RLC Digibridge
93