IET Labs, Inc. RLC Digibridge 1693 User And Service Manual page 116

Table of Contents

Advertisement

1693 RLC Digibridge
Character
Sequence
Status
1
Value of delta RLC or
ratio
2
Parameter
3
Normal or ratio mode
4
Units
5,6
7
Format
Sign
8
Number
9...15
16
17
Table 3-21: Data output format for RLC value, delta%, deltaRLC, or RATIO
* Underrange generally occurs when the measured impedance is lower than the nominal low limit (for automatic range change)
and there is a lower-impedance range. Overrange generally occurs when the measured impedance is higher than the nominal
high limit (for automatic range change) and there is a higher-impedance range. In either case, an appropriate range change
will improve accuracy. If there is no appropriate range to change to, the status is "range end ex tension "; see below. Range
limits are different if CONSTANT VOLTAGE OR CONSTANT CURRENT has been selected; see paragraph 3.5.6.
** Range 1 end extension if DUT impedance exceeds that of the largest R or L (smallest 0) in basic range 1. Range 4 end
extension if DUT impedance is less than that of the smallest R or L (largest 0) in basic range 4. No possibility of changing range
to improve accuracy.
*** If the measurement is outside of the display range (display goes blank), 9999999 is output in this numeric field.
92
Allowed
Purpose
Characters
(space)
U
O
E
I
(space)
-
/
L
C
R
G
Z
Y
(space)
-/
(space)H
mH
uF
nF
(space)S
mS
uS
(space)%
(space)Ω
(space)
(space)
-
0123456789.
(space)
(CR)
(LF)
Meaning
Normal Operation
Underrange held (reduced accuracy)*
Overrange held (reduced accuracy)*
End extension of range 1 or 1 (reduced accuracy)**
Invalid measurement due to signal overload
Normal display
Delta RLC Mode
Ratio Mode: Nominal Value/RLC Value
Inductance
Capacitance
Resistance
Conductance
Impedance
Admittance
Normal mode (one of 3 describe above)
Ratio Mode: RLC Value/Nominal Value
Henries
Millihenries
Microfarads
Nanofarads
Siemens
Millisiemens
Microsiements
Percent difference from nominal value
Ohms
kiloohms
Positive Value
Negative Value
Measured number, right justified in format field, like the
RLC display except the zero before the decimal point is
explicitly provided and this number can be as long as seven
characters. ***
The standard "carriage return" and "line feed" characters;
end of string.