Binning:
Pass bins: 13 pass bins for RLCGZY
Fail bins: 2 fail bins, RLCGZYΘ
Sorting capabilities:
Bin number, Delta RLC, Delta %, Value
DUT connections:
The 1693 has four BNC connections to make
four-terminal (Kelvin) connections to the device
under test. These connections are shielded and the
instrument ground has a guard for three-terminal
measurements. 1689-9602 1-meter BNC Cable
Set is provided. Various test fixtures and other
accessories are available.
Front of
1689-9602
Description
1693
BNC Cable
IL
Drive -
PL
Potential -
PH
Potential +
IH
Drive +
*High and Low Reversed for correct operation
Charged capacitor protection:
The instrument is protected from damage due to the
Stored Energy = ½CV
connection of charged capacitors with up to 1 joule
of stored energy:
Keyboard lock:
The keyboard can be locked to prevent inadvertent
changes to test conditions and sorting routines.
Parameter storage:
All keyboard settings and programmed nominal
values and bin limits can be stored in non-volatile
memory for automatic reentry upon power up .
Self-Check diagnostics:
Self-tests are performed during power-up to verify
proper operation and validity of calibration. Coded
error signals on the display notify the operator if a
problem is encountered.
Abbreviated Specifications
1689-9600 Remote
Test Fixture*
IL/Black Band
I+
PL/Black and
P+
White Band
PH/Red and
P-
White Band
IH/Red Band
I-
Joules
2
Special functions:
Several special features may be selected including:
• Direct Range Setting
• Range Extension (RATIO mode)
• Integration Time
• Blanking of lesser digits
• Signal reversal for hum removal under adverse
conditions
• A Median value routine
• Automatic parameter selection
• Quick acquisition routine
• Choice of data output on IEEE bus
• Interface Option: IEEE-488 Bus (1689-9640)
Zeroing:
Open and short circuit compensation
Power:
90-250 Vac
50-60 Hz
60 W max
Environmental conditions:
Operating conditions: 0° to +50°C, <85% RH
Storage conditions: -45°C to +75°C
Dimensions:
43.8 cm W x 14.3 cm H x 38.5 cm D (17.3" x 5.6"
x 15.2")
Weight:
6.4 kg (14 lbs)
Supplied accessories:
Power cable
1689-9602 bnc-to-bnc extender cable, 1-meter
Instruction manual
Calibration certificate traceable to SI
Optional accessories:
1700-03
Kelvin Test Leads
7000-04
Alligator Clip Leads, 1 Meter
7000-05
Chip Component Tweezers
1689-9602-2 bnc-bnc Extender Cable, 2-Meter
1689-9600
Remote Test Fixture
1689-9640
IEEE Digibridge Interface
1689-9604
Digibridge Calibration Kit
1689-9611
Rackmount Kit
1657-9600
Banana-Plug Extender Cable
1658-2450
Bias cable
Calibration certificate accredited to ISO17025
1693 RLC Digibridge