I. Determine the delta % reading as follows:
Select FUNCTION --- MEASURE
DISPLAY ---- DELTA %
J. Note the average reading. (For example:
range 1 may be - .8849).
K. Calculate the new K factor as shown in
Figure 5-14.
5. Store the new correction factor for each
range.
A. Select FUNCTION ---- ENTER
B. Select DISPLAY --- VALUE
C. Select R/Q --- until ohms LED is on
D. Enter the new K factor as a nominal value.
For example , range 1, press [-] [7] [9] [8] [.] [8] [3]
[=] [SHIFT] [NOM VALUE]
E. Store the K factor for each range:
for range 1 , press [1][6] [9] [3] [.] [ 1] [=]
[SHIFT] [SPECIAL] [SHIFT]
Service and Maintenance
Sample K factors:
Figure 5-14: Calculating the K factor
for range 2 , press [1][6] [9] [3] [.] [2] [=]
[SHIFT] [SPECIAL] [SHIFT]
for range 3 , press [1][6] [9] [3] [.] [3] [=]
[SHIFT] [SPECIAL] [SHIFT]
for range 4, press [1][6] [9] [3] [.] [4] [=]
[SHIFT] [SPECIAL] [SHIFT]
F. Select MEASURE RATE ---- MEDIUM to
FAST to SLOW
6. Once the K factors have been determined
and stored for all four ranges (step 4 through
5) repeat the recalibration procedure for all
ranges (see paragraphs 5.9.2 through 5.9.6).
Range 1:
-587
Range 2:
-39
Range 3:
-2.7
Range 4:
<±1
1693 RLC Digibridge
145

Advertisement