IET Labs, Inc. RLC Digibridge 1693 User And Service Manual page 111

Table of Contents

Advertisement

Parameter:
Value
9600
9600 BAUD
19200
19200 BAUD
38400
38400 BAUD
57600
57600 BAUD
115200
115200 BAUD
Equivalent legacy command:
none
Equivalent key:
none
:SYSTem:COMMunicate:SERial:STOPbits
Syntax:
:SYSTem:COMMunicate:SERial:STOPbits {1|2}
Description:
Set the number of RS-232 serial interface stop bits.
Parameter:
Value
1
1 stop bit
2
2 stop bits
Equivalent legacy command:
none
Equivalent key:
none
:SYSTem:COMMunicate:SERial:HANDshake
Syntax:
:SYSTem:COMMunicate:SERial:HANDshake
{NONe|XON|HARDware|MODem}
Description:
Set the RS-232 serial interface handshake.
Parameter:
Value
NONe
No handshake
XON
XOn/XOff software handshake
HARDware
Hardware handshake
MODem
Modem handshake
Equivalent legacy command:
none
Operation
Description
Description
Description
Equivalent key:
none
:SYSTem:ERRor[:NEXT]?
Syntax:
:SYSTem:ERRor[:NEXT]?
Description:
Reads the error message. (Query only)
Equivalent legacy command:
none
Equivalent key:
none
:SYSTem:VERSion?
Syntax:
:SYSTem:VERSion?
Description:
Reads the instrument version. (Query only)
Equivalent legacy command:
none
Equivalent key:
none
:STATus:OPERation[:EVENt]?
Syntax:
:STATus:OPERation[:EVENt]?
Description:
Reads the value of the Operation Status Event
Register. (Query Only)
Equivalent legacy command:
none
Equivalent key:
none