IET Labs, Inc. RLC Digibridge 1693 User And Service Manual page 20

Table of Contents

Advertisement

1693 RLC Digibridge
Kfv as a Function of Frequency and RMS Voltage:
12 Hz to
30 Hz to
100 Hz to
<30 Hz
<100 Hz
<250 Hz
1 V to 1.275 V AC Applied Voltage
5
3
0.25 V to < 1 V AC Applied Voltage
8
5
0.1 V to < 0.25 V AC Applied Voltage
12
8
0.03 V to < 0.1 V AC Applied Voltage
35
30
0.01 V to < 0.03 V AC Applied Voltage
90
80
* Multiply Kfv values by 5 from 3kHz to 20kHz for Range 1 (Z > 25 kohms)
** Not Specified
Measurement ranges and source impedance/
resistance:
The 1693 has 4 measurement ranges, Range 1 to
Range 4, that have specific source resistance. The
Ranges can be automatically or manually selected.
The source resistance does reduce the AC signal
applied to the DUT.
xviii
>1 kHz to
250 Hz
to
1 kHz
3 kHz
<1 kHz
2
1
0
1
3
2
1
2
6
5
4
5
25
20
14
15
70
60
50
50
Frequency
>3 kHz to
>6 kHz to
>10 kHz to
6 kHz*
10 kHz*
20 kHz*
1*
3*
5*
2*
4*
6*
6*
8*
10*
15*
20*
25*
50*
50*
60*
>20 kHz to
>20 kHz
200 kHz
50 kHz*
to 50 kHz*
8
18
70
12
25
80
20
50
100
40
100
**
70
**
**
Abbreviated Specifications