Equivalent key:
[MEASURE MODE]
:CONFigure:DISPlay
Syntax:
:CONFigure:DISPlay {BIN|DELTa|VALue|RLC}
Description:
Set measurement displayed on Digibridge front
panel.
Parameter:
Value
BIN
Display Bin #
DELTa
Display Delta %
VALue
Display Value
DRLC
Display Delta RLC
Equivalent legacy command:
D0, D1, D2, D3
Equivalent key:
[DISPLAY]
:CONFigure:DISPLay:RATio
Syntax:
:CONFigure:DISPlay:RATio
{NOMValue|VALNominal}
Description:
Set Digibridge front panel to display measurement
ratio.
Parameter:
Value
NOMValue
Display Ratio: nominal value / RLC value
VALNominal
Display Ratio: RLC value / nominal value
Equivalent legacy command:
D4, D5
Equivalent key:
[DISPLAY] + nominal value
Operation
Description
Description
:CONFigure:DISPLay:BINRatio
Syntax:
:CONFigure:DISPLay:BINRatio
{NOMValue|VALNominal}
Description:
Set Digibridge front panel to display measurement
bin on ratio.
Parameter:
Value
NOMValue
Display Bin on ratio: nominal value / RLC value
VALNominal
Display Bin on ratio: RLC value / nominal value
Equivalent legacy command:
D6, D7
Equivalent key:
[DISPLAY] + nominal value
:CONFigure:FUNCtion
Syntax:
:CONFigure:FUNCtion {MEASure|ENTer}
Description:
Set function mode to measurement or setting entry.
This command affects user keyboard entry.
Parameter:
Value
MEASure
Set function to Measure
ENTer
Set function to Enter (enable setting configuration)
Equivalent legacy command:
P0, P1
Equivalent key:
[MEASURE] key, current state indicated by
MEASURE and ENTER LEDs.
:CONFigure:RANGe
Syntax:
:CONFigure:RANGe {HOLD|1|2|3|4|AUTo}
Description:
Set the measurement range.
1693 RLC Digibridge
Description
Description
81

Advertisement

Table of Contents
loading

Related Products for IET Labs, Inc. RLC Digibridge 1693

Table of Contents