Troubleshooting; Alarm - Cause - Remedy - Dräger Isolette 8000 plus Instructions For Use Manual

Hide thumbs Also See for Isolette 8000 plus:
Table of Contents

Advertisement

11

Troubleshooting

11.1
Alarm – Cause – Remedy
The Isolette 8000 plus displays alarm messages in the Trend/Alarm window. If two
or more system alarms occur simultaneously, or one after the other, the alarm
messages of the highest priority (i.e., Controller Failure alarms) are presented first.
They are followed by all other messages in sequence. A total of 6 messages can be
presented in the Trend/Alarm window.
In addition to a displayed message, alarm conditions are also indicated by an
acoustic alarm signal and a lighted LED.
In the following table, the alarm messages are listed in alphabetical order. If an
alarm occurs, the table helps to identify causes and remedies. The different causes
and remedies should be worked through in the order listed until the alarm has been
resolved.
To help service personnel, all failures and error codes should be recorded.
11.1.1
System alarms
Alarm
Alarm
priority
Medium
Check Settings
Medium
Controller Failure 1-14
(For Controller Failure alarms 3, 4, and
7, alarm is delayed for 1 minute after
the alarm condition occurs)
Medium
Controller Failure 15-17
Medium
Heater Failed 1
(Alarm can be silenced for 5 minutes)
Instructions for use Isolette 8000 plus SW 5.n
Cause
A failure is detected with the
controller non-volatile mem-
ory test.
An internal malfunction
occurred.
This is the first time the
device was installed or new
software was loaded onto
the controller.
Controller memory failure
was detected.
The heater thermocouple
voltage exceeds 40 mV.
Troubleshooting
Remedy
Check all settings to make
sure that they are config-
ured correctly.
Switch off the incubator and
then turn it on again. If the
failure continues, switch off
the incubator and remove it
from service.
Press the Audio
Paused/Reset key.
Reset settings to factory
defaults by pressing the
Hard Default button in the
system configuration menu.
Check the desired settings.
If the failure continues,
switch off the incubator and
remove it from service.
Switch off the incubator, and
remove it from service.
115