Dräger Isolette 8000 plus Instructions For Use Manual page 76

Hide thumbs Also See for Isolette 8000 plus:
Table of Contents

Advertisement

Getting started
1 Ensure that the bed is level and not in the Trendelenburg or Reverse
Trendelenburg position.
 The bubble on the scale level indicator is within the circle (see Symbols on
2 Select display 2, and press the Weight button.
 The weight screen is displayed.
3 If necessary, remove any objects from the bed.
4 Press the ->0/T<- button twice (see Buttons on page 42).
 The weight display reads zero and the background color changes.
5 Place a weight of known value, but less than 7 kg (15 lb), on the bed and verify
that the correct weight is displayed.
6 Remove the weight from the bed, and press the Home button.
7.2.9
Utility shelf (optional component) functional check
1 Ensure that the utility shelf and support arm are secure.
2 Ensure that the utility shelf rotates with the desired resistance.