8.7.3
Starting kangaroo mode:
1 If necessary, unlock the keypad.
2 At display 2, press the Kangaroo button (A) to select the kangaroo mode.
The kangaroo mode screen is shown
The kangaroo screen shows the rotating wheel (B).
The duration of kangaroo mode is continuously shown on the elapsed time
34.5
36.1
40
35
30
25
20
3 Open the hood or access panel and remove the patient from the incubator.
4 Place the patient on the parent's naked chest.
5 To prevent heat loss, cover the patient with a blanket.
6 If skin temperature probes are used, ensure that they are connected correctly to
both the device and to the patient.
In kangaroo mode, the yellow probe monitors the central skin temperature.
However, the central skin temperature value does not affect the temperature in
the patient compartment.
7 Close the hood or access panel. Ensure that the hoses and cables do not get
caught or kinked.
Instructions for use Isolette 8000 plus SW 5.n
meter (C).
35.0
C
1.8
43
21
A
B
99:58
50
21
Operation
99