Assembly and preparation
10 Reinstall the sensor module in the hood, and position it in the normal (innermost)
position.
11 Reconnect the cables that were previously removed.
 Sensor module cable is connected to the shell.
 Scale cable and temperature probes cables are connected to the sensor
12 Switch on the incubator at the controller on/off switch, while pressing the Audio
Paused/Reset key.
 System configuration mode appears.
13 Confirm that the Altitude field contains the correct height above sea level of the
site, to the nearest 1000 feet (300 meters). Consult a map or Google Earth if the
height above sea level is not known. Correct if necessary.
14 Note the oxygen calibration level.
 Oxygen calibration level is either 21% or 100%.
15 If the oxygen calibration level is 100%, the 100% calibration fixture (with hose
barb) must be installed below the sensor module. If the oxygen Calibration Level
is 21%, the slide lock fixture (without hose barb) must be installed below the
sensor module.
16 Switch off the incubator at the controller on/off switch, wait 1 minute, and turn
the incubator back on in normal mode.
17 When 30 minutes have elapsed since step 1 (exposing the oxygen sensors to
air), begin oxygen calibration (see Oxygen sensor calibration (option) on page
160). Note carefully the use of room air (21% oxygen) or pure oxygen, as
dictated by the presence or absence of the 100% oxygen calibration fixture on
the hood.
56
G
E
F
module.
Instructions for use Isolette 8000 plus SW 5.n