8.5.1
Displaying trends
If no keys are pressed within 15 seconds of making any selection, the screen
automatically reverts to the previous display.
1 If necessary, unlock the keypad.
2 At display 1, press the Display Selection key.
Display 2 appears.
3 At display 2, press the Trend button to select the Trend display.
Trend menu items appear.
4 Press and hold the Clear button (A).
All Trend displays are cleared.
5 To select the trend period: 2, 4, 8, 12, or 24 hours, press the Hours button (B).
The Hours button is not applicable for trends of the infant weight. The trend
interval for infant weight is 7 days.
6 To select one of the following Trend displays, press the Display button (C)
repeatedly until the desired setting is shown:
Oxygen (optional)
Humidity (optional)
Weight (optional)
Instructions for use Isolette 8000 plus SW 5.n
40
35
30
25
20
Air temp
Skin temp
Heater power %
Operation
B
A
C
D
E
93