Dräger Isolette 8000 plus Instructions For Use Manual page 154

Hide thumbs Also See for Isolette 8000 plus:
Table of Contents

Advertisement

Service
17 Slide the new collection bottle onto the bracket (E) located underneath the left
side of the incubator.
18 Remove the caps from the ports on the collection bottle.
The large port on top of the collection bottle must remain uncapped during
operation to ensure that water drains properly into the bottle.
19 Attach the clamp to the hose (F).
20 Connect the condensation management hose to the patient port (G) on the side
of the collection bottle.
21 Install the heater/impeller cover.
22 Install the upper cover.
23 Install the T-bars on the upper cover.
24 Install the bed, X-ray tray, and scale (optional).
25 Install the mattress.
26 Close the hood.
13.5.1.4
Collection bottle replacement (if installed)
CAUTION
Risk of injury or damage to the device
Excess fluid in the collection bottle can lead to spillage and a safety hazard
► To avoid overflow, check the water level in the collection bottle every 8 hours.
► When the water level reaches the maximum mark on the bottle, replace the
bottle.
154