Dräger Isolette 8000 plus Instructions For Use Manual page 130

Hide thumbs Also See for Isolette 8000 plus:
Table of Contents

Advertisement

Reprocessing
12.1.2
Removing the bed and related components, and optional scale
1 Disconnect the cables from the sensor module.
2 Slowly raise the hood (A).
3 Remove the mattress (B).
4 If the incubator is equipped with a weighing scale, perform steps 5 through 7;
otherwise go to step 8.
5 Disconnect the scale cable from the sensor module.
6 Remove the cable from the clamps that hold it to the incubator.
7 Lift the scale (C) from the bed.
8 Remove the bed (D) and X-ray tray (inside bed).
9 Remove the T-bars (E).
10 Remove the upper cover (F).
130
B
C
D
E
E
F
Instructions for use Isolette 8000 plus SW 5.n
A

Hide quick links: