Problem/
alarm condition
11.2.2
No data is displayed
Problem/
alarm condition
No data is displayed on screen at
system start-up (only
static/noise).
11.2.3
Adjustment of device height not possible
Problem/
alarm condition
The height adjustment trolley
does not move up or down.
11.2.4
Water is leaking
Problem/
alarm condition
Water is leaking onto the floor
Instructions for use Isolette 8000 plus SW 5.n
Cause
Over-current protection fuses in
the height adjustment trolley
have blown.
Cause
Internal controller failure.
Cause
The main on/off switch is OFF.
Over-current protection fuses in
the height adjustment trolley
have blown.
The height adjustment trolley
was operated for more than
3 minutes continuously which
could result in the height adjust-
ment trolley motor overheating.
Cause
Condensation management sys-
tem is not installed.
Condensation management hose
is incorrectly installed:
Drain plug on hose is not prop-
erly installed.
Hose is connected to the wrong
port on the collection bottle.
Troubleshooting
Remedy
Switch off the incubator and
replace the fuses (see Fuse
replacement on page 151).
Remedy
Switch off the incubator and
remove it from service.
Remedy
Switch on the main on/off switch
on the trolley.
Switch off the incubator and
replace the fuses (see Fuse
replacement on page 151).
Wait 10 minutes and try to oper-
ate the trolley again.
Remedy
Install condensation manage-
ment system (see Condensation
management system replace-
ment (if installed) on page 152).
Ensure that the drain plug is fully
inserted into the drain opening so
that it is flush with the heater well
floor.
Ensure that the hose is con-
nected to the patient port on the
side of the collection bottle.