Calibration of the oxygen system is dependent on the altitude.
13.7.1.1
Oxygen sensor calibration to room air (21%)
WARNING
Risk of death or serious injury
If the 100% oxygen calibration fixture is used during the 21% calibration procedure,
the calibration will be inaccurate.
► To perform the 21% oxygen calibration procedure, ensure that the incubator is
equipped with the standard sensor module slide-lock.
To calibrate the oxygen sensor to room air (21%):
1 Verify the oxygen calibration level setting (21% or 100%) at the system
configuration menu (Refer to System configuration menu on page 113).
2 If necessary, configure the oxygen calibration level setting for 21%.
The proper oxygen calibration altitude is selected in the system configuration
3 Ensure that the incubator is equipped with the standard sensor module slide-
lock, not the 100% oxygen calibration fixture.
4 At display 1, press the Oxygen button.
5 Press the On button, and then the Cal button.
The message Slide Out Sensor appears.
6 Within 5 seconds, withdraw the sensor module from the hood to prevent a
Cal Fail message. To withdraw the module, pull down the lock (A) and slide out
the module (B) from the hood until it stops.
When calibration is complete, the message Cal Pass is displayed in the
7 Slide the sensor module inside the hood, and then press the On button.
8 If the message Cal Fail is displayed, refer to In case of calibration failure on
page 162.
9 If the calibration procedure is unsuccessful a second time, refer the device to
service personnel.
Instructions for use Isolette 8000 plus SW 5.n
menu (Refer to System configuration menu on page 113).
A
oxygen window.
B
Service