Dräger Isolette 8000 plus Instructions For Use Manual page 79

Hide thumbs Also See for Isolette 8000 plus:
Table of Contents

Advertisement

8.1.1.1
Activating temperature override mode when setting the temperature
1 Press the Up Arrow key (E) and raise the set temperature to 37.0 °C (98.6 °F).
G
2 To activate the temperature override mode, press the >37°C button (F).
 The >37°C LED indicator (G) illuminates and the color of the >37°C button
The >37°C button is inoperative until the air set temperature is set to 37.0 °C
(98.6 °F).
3 To raise the set temperature beyond 37.0 °C (98.6 °F) up to 39.0 °C (102.2 °F),
press the Up Arrow key (E).
 Pressing this key raises the temperature in 0.1° increments.
8.1.2
Skin temperature mode
The incubator must be pre-warmed for at least 1 hour in air temperature mode
before switching to skin temperature mode.
To select the set point in skin temperature mode:
1 If necessary, unlock the keypad.
If no keys are pressed within 15 seconds of making any selection, the screen
automatically reverts to the previous display.
Instructions for use Isolette 8000 plus SW 5.n
40
35
30
25
20
label changes.
F
E
Operation
79