Defaults
Air temperature upwards deviation limit
default
Air temperature downwards deviation
limit default
Skin temperature deviation limit default
Oxygen deviation limit default
8.8.1
Entering the setup menu
1 If necessary, unlock the keypad.
2 At display 2, press the Setup button (A).
 The Temp/O2 setup screen appears.
21.9
35.3
Setup Temp/O2
Air Temp Upper Deviation +
Air Temp Lower Deviation -
Skin Temp Deviation +/-
Oxygen Deviation +/-
Instructions for use Isolette 8000 plus SW 5.n
+1.5 °C (+2.7 °F)
- 2.5 °C (-4.5°F)
1.0 °C (1.8 °F)
3%
A
35.0
B
22.2
2.1
2.4
0.7
3
Operation
C / F
101