Dräger Isolette 8000 plus Instructions For Use Manual page 69

Hide thumbs Also See for Isolette 8000 plus:
Table of Contents

Advertisement

2 Rotate the knob clockwise until it stops.
3 Rotate the left bed-tilt knob clockwise until it stops.
4 Rotate the knob counterclockwise until it stops.
7 Check the bed operation:
1 Rotate the locking knobs, and open the front access panel.
2 Pivot the front access panel to the full open position (hanging straight down).
3 Slide out the bed to the fully extended position (C).
4 Carefully lean on the bed to ensure that it is properly supported and provides
5 Return the bed.
6 Close the front access panel, and rotate both locking knobs until they are
Instructions for use Isolette 8000 plus SW 5.n
 The bed is level.
B
 The bed is in the full "up" position.
 The bed is level.
a firm infant platform.
fully engaged.
C
A
Getting started
69