128
Cause
This message is displayed during
the zeroing and calibration rou-
tine.
This message is displayed when
there is too little weight on the
bed during calibration (for exam-
ple, no mattress)
This message is displayed when
an extra weight on the bed
exceeds 4000 g when zeroing
during infant weight.
Instructions for use Isolette 8000 plus SW 5.n
Remedy
Information only, no action
required.
Replace mattress and recali-
brate.
Remove extra weight and retry
zeroing. If it fails again, remove
scale from service.