Instructions For Use (Qualified Personnel) - Otto Bock Genium Protector 4X880 Series Instructions For Use Manual

Hide thumbs Also See for Genium Protector 4X880 Series:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 10
Observe the following sections in these instructions for use.
CAUTION
Improper handling, assembly and disassembly
Injury due to faulty operation or malfunction of the product.
Instruct the patient in the proper disassembly, assembly and handling of the product.
INFORMATION
When the protector is installed, the battery charger cable has to point to the upper closure. A
correct knee joint charging process is only ensured with this alignment.
5 Scope of Delivery and Accessories
Scope of Delivery
Quant­
Product
ity
1 pc.
Genium Protector
1 pc.
Upper closure
1 pc.
Lengthwise closure
1 pc.
Foot cuff
1 pc.

Instructions for use (qualified personnel)

1 pc.
Instructions for use (user)
Accessories
Quant­
Product
ity
-
Lengthwise closure
-
Foot cuff
6 Preparation for Use
Required Tools
Designation
Vibrating/band saw
Knife
Loctite® 241 Thread Locking Compound
Crepe Adhesive Tape
Torque Wrench
Tape Measure
6.1 Adapting the Protector
6.1.1 Shortening the Protector – Preliminary Shortening
1) Remove the prosthesis from the residual limb.
2) Remove the shoe from the prosthetic foot.
3) Loosen the foot connection cap of the prosthetic foot.
4) Check whether the closures on the protector are properly fastened and the protector is not
twisted.
Article number
4X880=*
-
-
4P880=*
647G942
646D298
Article number
4P880=R
4P880=*
Reference number
-
-
636K13
627B6=15
-
-
Ottobock | 13