STEP 3. Install the Network Analyzer Modules

STEP 3. Install the Network Analyzer Modules
These modules can be used in a chassis with a PXIe, or PXI-H chassis slots.
NOTE
Recommended Best Practices to Ensure Proper and Safe Module Operating
Conditions
Figure 1-2
M9485A chassis air flow

Tools Required for the Installation Procedure

Ensure proper chassis air flow is maintained.
The use of a Keysight M9018A chassis is recommended for optimum
temperature control. If a different chassis is used, make sure it provides
thermal protection if fans become inoperable or forced air cooling is
obstructed.
— If using the Keysight M9018A PCIe Cable Interface, download the
Keysight Multiple PXIe and AXIe Chassis System Configuration Tool.
Save the file to your local hard drive; do not try to open the file from the
Internet location.
Use slot blockers and EMC filler panels in empty module slots to ensure
proper operating temperatures. Keysight chassis and slot blockers optimize
module temperature performance and reliability of test.
Set chassis fans to high or auto. Do not disable fans.
Position chassis to allow plenty of space around chassis air intake and fan
exhaust.
At ambient temperatures above 45° C (113° F) set the chassis fan to high.
Phillips #1 torque driver, set to 4 in-lb (0.45 N.m) - not supplied.