Installation Procedure - Keysight Technologies M9485A Startup Manual

Pxie multiport vector network analyzer
Table of Contents

Advertisement

Getting Started
STEP 3. Install the Network Analyzer Modules

Installation Procedure

PXI hardware does not support "hot-swap" (changing network analyzer modules while
CAUTION
power is applied to the chassis) capabilities. Before installing or removing a module
to/from the chassis, power-off the chassis to prevent damage to the module.
1. Make sure that the line cord is plugged in to establish earth ground and that
2. Wear a grounded wrist strap for this procedure (see callout 2 in
3. If the chassis has multiple fan speed settings, ensure that the fan switch is
4. Position the chassis to provide ample space between the chassis fan intake
5. Before inserting the network analyzer module into the chassis, back-out the
6. Holding the module by the injector/ejector handle, slide it into a slot, as
Socket adapter (part number M9485-23001) - supplied.
Torque driver (for use with socket adapter), set to 8 in-lb (0.91 N.m) - not
supplied.
Cable removal tool (part number 5002-3361) - supplied.
Torque Wrench for SMA, set to 8 in-lb (0.91 N.m) - not supplied.
5.5 mm open end wrench - not supplied.
The network analyzer module can be installed in any PXIe or hybrid PXI
slot marked with a peripheral slot compatibility image (solid black
circle fro PXIe, or solid black circle with the letter "H" for hybrid.
the chassis power switch is Off (see callouts 1 and 3 in
page
11).
on page
11).
set to high or auto.
and exhaust vents. Blockage by walls or obstructions affects the air flow
needed for cooling. (Refer to the chassis documentation for more
information about cooling.)
top mounting screw (see callout 7 in
to ensure that there is no interference between the screws and the
mounting rails.
shown in callouts 4 – 6 in
a. Install the module into the slot of the chassis by placing the module card
edges into the front module guides (top and bottom).
Figure 1-3 on page
Figure 1-3 on page
11.
10
Figure 1-3 on
Figure 1-3
11) on the module