Thermo Scientific Vanquish VC-D40 Operating Manual page 69

Variable wavelength detectors
Table of Contents

Advertisement

Parts required
Tools required
Preparations
Follow these steps
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
In addition, observe the guidelines for proper connection of the flow cell
in
General Information and Guidelines
Inlet capillary
Detector waste line
For instructions on connecting the waste line, follow the steps in
Connecting the Detector Waste Line
For the detector waste line: Tubing cutter (optional)
TIP For easier guiding of capillaries to the flow cell inlet, remove the
right door of the detector, see
1. Flush the system modules and capillaries upstream of the device to
waste before you connect the flow cell to the system flow path.
Refer to the Vanquish System Operating Manual.
2. On the flow cell, remove the plugs from the flow cell inlet and
outlet.
TIP Store the plugs of the flow cell, for example in the flow cell
packaging, to have them easily available when storing or shipping the
flow cell.
3. Install the partition panel plugs (if not present yet). See
Partition Panel Plugs
1. Connect the inlet capillary to the flow cell.
2. Connect the detector waste line to the flow cell.
See also
2
Connecting the Inlet Capillary (} page 70)
2
Connecting the Detector Waste Line (} page 71)
(} page 62).
(} page 71).
Replacing the Doors
(} page 65).
5
Installation
(} page 126).
Installing the
Page 69