Variable wavelength detectors
Table of Contents

Advertisement

Type
Noise (Multi-channel mode)
Drift
Linearity
USB communication
I/O Interface
System Interlink
Analog output
Control
Materials in the flow path
Safety features
Good Laboratory Practice (GLP)
features
Temperature reference conditions for noise, drift and linearity specifications
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
Specification
<± 10.0 µAU (typical: <± 7.0 µAU), wavelengths: 254 nm and 280 nm, only
deuterium lamp turned on
<± 10.0 µAU (typical: <± 7.0 µAU), wavelengths: 450 nm and 520 nm, only
tungsten lamp turned on
Standard flow cell with 1 mL/min HPLC-grade water, degassed, column in
flow path, time constant: 2 seconds, data collection rate: 1 Hz
All values reached within 60 minutes of warm-up (typical).
< 0.1 mAU/hour, wavelength: 254 nm, only deuterium lamp turned on
< 0.1 mAU/hour, wavelength: 520 nm, only tungsten lamp turned on
Standard flow cell with 1 mL/min HPLC-grade water, degassed, column in
flow path, time constant: 2 seconds
All values reached within 60 minutes of warm-up (typical).
< 5% at 2.5 AU (ASTM), wavelength: 272 nm, component: caffeine, only
deuterium lamp turned on
Standard flow cell
1 USB port (USB 2.0, "B" type connector)
1 USB hub with 3 ports (USB 2.0, "A" type connectors)
2 digital I/O ports
2 System Interlink ports (RJ45-8 connectors)
2 analog outputs via optional plug-in expansion board to output absorption
channels
Resolution: 20 bit
Maximum data rate: 50 Hz
Outputs can be configured via software (output voltage range 0 to 1 V or
0 to 10 V, sensitivity and offset)
Chromeleon 7
The device can be operated also with other data systems. For details,
contact the Thermo Fisher Scientific sales organization.
Keypad with 3 buttons for performing certain functions directly from the
device
See the Specifications for the flow cells.
NOTICE For information about the chemical resistance of materials refer to
the technical literature.
Power-up check of optics, cooling fans, motors and electronics
Leak detection and safe leak handling
Predictive Performance functions for scheduling maintenance procedures
based on the actual operating and usage conditions of the device. This
includes monitoring of lamp age, and lamp ignitions for UV and VIS lamp.
All system parameters are logged in the Chromeleon Audit Trail.
For drift tests, ASTM requires that temperature changes stay below 2 °C/
hour over a period of one hour. The drift specification above is based on
these conditions. Larger changes in the ambient temperature will result
in a larger drift.
9
Specifications
Page 151