7.12.2 Transporting the Device to a New Location

Preparations
Follow these steps
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
NOTICE
The flow cell opening on the device is sensitive to dust and debris.
If no flow cell is installed to the detector, install a diagnostic cell to the
flow cell opening of the detector.
5. Remove the device or slide-in module from the system stack as
required:
Install the carrying handles and remove the device from the
Vanquish system. Follow the instructions on dismounting the system
stack in the Transporting or Shipping the System section of the
Vanquish System Operating Manual.
‑or‑
Remove the slide-in module from the device enclosure in the system
stack (see
Removing the Slide-In Module
See also
2
Removing the Flow Cell (} page 116)
2
Installing a Diagnostic Cell (} page 118)
Prepare the device for transport. See
(} page 128).
1. Observe the notes for handling and lifting the device safely.
2. Transport the device to the new location.
3. Install and set up the device in the system stack. Follow the
instructions on mounting the system stack in the Vanquish System
Operating Manual.
4. Set up the device:
a) Connect the device and set up flow connections (see
(} page 45)).
b) Prepare the device for operation (see
Operation
(} page 82)).
5. Before starting an analysis, let the device equilibrate and be sure
that it is ready for operation.
7
Maintenance and Service
(} page 131)).
Preparing the Device for Transport
Preparing the Device for
Installation
Page 129