Set the Pressure Module to Zero

Use this option to write a new zero pressure value to the pressure
module in use. The sensor adjustment is permitted if it obeys the
condition that follows:
Adjustment ≤10% FS positive pressure value (for the Sensor).
2.3.3

Error Indications

Under range:
Over range:
If the display shows <<<< (under range) or >>>> (over range):
Make sure that the range is correct.
Make sure that all the related equipment and connections are
serviceable.
Page 51 of 79
3. Use
to start the Leak Test
4. Use
to stop the Leak Test.
Note: To make a temporary adjustment for zero, use the
Tare function.
The display shows this symbol for this condition:
Reading < 110% Negative Full Scale (Pressure)
Reading < 102% Negative Full Scale (Electrical)
The display shows this symbol for this condition:
Reading > 110% Positive Full Scale (Pressure)
Reading > 102% Positive Full Scale (Electrical)
<<<<<
>>>>>