User Manual
2.3.2
Set the Pressure Module to Zero
Use this option to write a new zero pressure value to the pressure
module in use. The sensor adjustment is permitted if it obeys the
condition that follows:
Adjustment ≤10% FS positive pressure value (for the Sensor).
•
2.3.3
Error Indications
Under range:
Over range:
If the display shows <<<< (under range) or >>>> (over range):
Make sure that the range is correct.
•
Make sure that all the related equipment and connections are
•
serviceable.
Page 51 of 79
3. Use
to start the Leak Test
4. Use
to stop the Leak Test.
Note: To make a temporary adjustment for zero, use the
Tare function.
The display shows this symbol for this condition:
Reading < 110% Negative Full Scale (Pressure)
Reading < 102% Negative Full Scale (Electrical)
The display shows this symbol for this condition:
Reading > 110% Positive Full Scale (Pressure)
Reading > 102% Positive Full Scale (Electrical)
<<<<<
>>>>>
[EN] English – K0571 Issue 1
Druck DPI611