Set The Pressure Module To Zero; Example Procedure: Measure Pressure - GE Druck DPI 620-IS User Manual

Advanced modular calibrator
Table of Contents

Advertisement

4.4.3 Set the pressure module to zero

TAP the window
(Maximise)
1
See the new reading

4.4.4 Example procedure: Measure pressure

1
2
[EN] English - K0460
Use this option to write a new zero pressure value to the
pressure module.
The sensor adjustment is permitted if it obeys this condition:
Adjustment ≤ 10% FS positive pressure value
Note: To make a temporary adjustment for zero, use the Tare
function; see
Section
Select Settings
2
5
Go back to the Home window.
3
P1
(for the Sensor)
2.9.1.
Accept the adjustment value
Select Zero
3
p1
p1
p2
p2
P2
Pressure indicator operation (MC 620-IS) 4-5
Issue 1
4