Procedures (Mv Or Volts Input); Procedures (Ma Input) - GE Druck DPI 832 User Manual

Electrical loop calibrator
Hide thumbs Also See for Druck DPI 832:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Selection sequence:
➤ Select Task (Table 2) ➤ Set Up (Table 1) ➤ Calibration ➤.
Display: Enter
1
Calibration PIN
■ ■
Factory PIN = 4321
Table 9: Calibration options
Options
Description
...
To calibrate the specified input: mV, Volts, mA
...
To calibrate the specified output: mA, mV, Volts
IDOS ...
UMM only. To calibrate the specified IDOS UMM.
Refer to the user manual for the IDOS UMM.
Calibration Due: To set the date of the next
calibration for the instrument.
After the specified calibration date, there is a
warning message. There is a selection box to stop
the warning.
To change the calibration PIN (Personal
Identification Number).
When you select a channel, the display shows the
applicable instructions to complete the calibration.
When the calibration is complete, select Calibration Due
and set the new calibration date for the instrument.
Calibration - Procedures (mV or Volts input)
1. Connect the instrument to the calibration equipment
(Figure 1).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a three-point
calibration (-FS, Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable Volts input task and range (Table 2).
5. Then apply the input values that are applicable to the
calibration:
• mV: -120, -60, -30, 0 (short circuit)
Then mV: 0, 30, 60, 120
OR
• Volts (V): -30, -15, -5, 0 (short circuit)
Then volts (V): 0, 5, 15, 30
10
- [EN] English
Menu:
2
3
Select Channel
+
(Table 9)
6. Make sure that the error is in the specified limits
(Table 10 or Table 11).
4
5
Applied
mV
±120
±60
±30
0 (Short circuit)
Applied
V
±30
±15
±5
0 (Short circuit)
Calibration - Procedures (mA input)
1. Connect the instrument to the calibration equipment
(Figure 4).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a three-point
calibration (-FS, Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable mA input task (Table 2) and apply these
values:
• mA: -55, -40, -24, -18, -12, -6, 0 (open circuit)
Then mA: 0, 6, 12, 18, 24, 40, 55.
5. Make sure that the error is in the specified limits
(Table 12).
Applied
mA
±55
±40
±24
±18
±12
±6
0 (open circuit)
Table 10: mV input error limits
Calibrator
error
(mV)
0.0013
0.0008
0.0006
-
Table 11: Volts (V) input error limits
Calibrator
error
(V)
0.00058
0.00011
0.00006
-
Table 12: mA input error limits
Calibrator
error
(mA)
0.0022
0.0018
0.0014
0.0004
0.0003
0.0002
-
Permitted
DPI 832 error
(mV)
0.03
0.02
0.02
0.01
Permitted
DPI 832 error
(V)
0.004
0.002
0.001
0.001
Permitted
DPI 832 error
(mA)
0.005
0.004
0.003
0.003
0.002
0.002
0.001
K393 Issue 3